If process variables continuous reading is enabled, a warning message will be displayed. To continue, un-check the continuous reading in the field Continuous reading in the process variables window.
Calibration eliminates errors e.g. sensor and whole computing track aging errors.
You should follow the wizard instructions.
After entering electric current value for lower calibration point in the field Calibration point, choose the button Next.
The transmitter will produce the adjusted current value. The transmitter can reject the adjusted value; this will indicate that the value is too high or too low.
Example: For a transmitter with 4...20mA output, the value must equal 4 mA
You should follow the wizard instructions.
When miliammeter readings are stabilized and entered into the field Measured current value, choose the Send button.
The transmitter can reject the value. This can indicate incorrect configuration, or improper transmitter / ammeter operation.
You should follow the wizard instructions.
After entering electric current value for upper calibration point in the field Calibration point, choose the button Next.
The transmitter will produce the adjusted current value. The transmitter can reject the adjusted value; this will indicate that the value is too high or too low.
Example: For a transmitter with 4...20mA output, the value must equal 20 mA
You should follow the wizard instructions.
When miliammeter readings are stabilized and entered into the field Measured current value, choose the Send button
The transmitter can reject the adjusted value. This can indicate incorrect configuration, no zeroing, or improper transmitter / ammeter operation.
Note: Cancelling calibration is possible with the command Cancel analog output calibration in Transmitter parameters window